User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Power Designs 2005:Sub 1 DATE: 30-Jul-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 12 NUMBER OF LINES: 105 CONFIGURATION: HP 34401A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R N P F W 1.003 HEAD {Vernier} 1.004 HEAD * CALIBRATION ACCURACY TEST * 1.005 JMP 2.001 1.006 EVAL 2.001 DISP Set the UUT front panel controls as follows: 2.001 DISP 100µV/DIV .................................... 0 2.002 34401 mV N 2W 2.003 MEME >1 2.004 DISP Set the UUT front panel controls as follows: 2.004 DISP 100µ/DIV ................................. .1mV 2.005 34401 mV N 2W 2.006 MATH mem = mem - m[1] 2.007 ACC 100 .100mV .005% .0035/ 2.008 MEME 2.009 MEMC mV 0.1% 1U #! Test Tol 1.0001, Sys Tol 0.0001007, TUR 9931.480 (>= 4.00). 3.001 DISP Set the UUT front panel controls as follows: 3.001 DISP 100µ/DIV ................................. .2mV 3.002 34401 mV N 2W 3.003 MATH mem = mem - m[1] 3.004 ACC 100 .200mV .005% .0035/ 3.005 MEME 3.006 MEMC mV 0.1% 1U #! Test Tol 1.0002, Sys Tol 0.0001014, TUR 9863.905 (>= 4.00). 4.001 DISP Set the UUT front panel controls as follows: 4.001 DISP 100µ/DIV ................................. .3mV 4.002 34401 mV N 2W 4.003 MATH mem = mem - m[1] 4.004 ACC 100 .300mV .005% .0035/ 4.005 MEME 4.006 MEMC mV 0.1% 1U #! Test Tol 1.0003, Sys Tol 0.0001021, TUR 9797.258 (>= 4.00). 5.001 DISP Set the UUT front panel controls as follows: 5.001 DISP 100µ/DIV ................................. .4mV 5.002 34401 mV N 2W 5.003 MATH mem = mem - m[1] 5.004 ACC 100 .400mV .005% .0035/ 5.005 MEME 5.006 MEMC mV 0.1% 1U #! Test Tol 1.0004, Sys Tol 0.0001028, TUR 9731.518 (>= 4.00). 6.001 DISP Set the UUT front panel controls as follows: 6.001 DISP 100µ/DIV ................................. .5mV 6.002 34401 mV N 2W 6.003 MATH mem = mem - m[1] 6.004 ACC 100 .500mV .005% .0035/ 6.005 MEME 6.006 MEMC mV 0.1% 1U #! Test Tol 1.0005, Sys Tol 0.0001035, TUR 9666.667 (>= 4.00). 7.001 DISP Set the UUT front panel controls as follows: 7.001 DISP 100µ/DIV ................................. .6mV 7.002 34401 mV N 2W 7.003 MATH mem = mem - m[1] 7.004 ACC 100 .600mV .005% .0035/ 7.005 MEME 7.006 MEMC mV 0.1% 1U #! Test Tol 1.0006, Sys Tol 0.0001042, TUR 9602.687 (>= 4.00). 8.001 DISP Set the UUT front panel controls as follows: 8.001 DISP 100µ/DIV ................................. .7mV 8.002 34401 mV N 2W 8.003 MATH mem = mem - m[1] 8.004 ACC 100 .700mV .005% .0035/ 8.005 MEME 8.006 MEMC mV 0.1% 1U #! Test Tol 1.0007, Sys Tol 0.0001049, TUR 9539.561 (>= 4.00). 9.001 DISP Set the UUT front panel controls as follows: 9.001 DISP 100µ/DIV ................................. .8mV 9.002 34401 mV N 2W 9.003 MATH mem = mem - m[1] 9.004 ACC 100 .800mV .005% .0035/ 9.005 MEME 9.006 MEMC mV 0.1% 1U #! Test Tol 1.0008, Sys Tol 0.0001056, TUR 9477.273 (>= 4.00). 10.001 DISP Set the UUT front panel controls as follows: 10.001 DISP 100µ/DIV ................................. .9mV 10.002 34401 mV N 2W 10.003 MATH mem = mem - m[1] 10.004 ACC 100 .900mV .005% .0035/ 10.005 MEME 10.006 MEMC mV 0.1% 1U #! Test Tol 1.0009, Sys Tol 0.0001063, TUR 9415.804 (>= 4.00). 11.001 DISP Set the UUT front panel controls as follows: 11.001 DISP 100µ/DIV .................................. 1mV 11.002 34401 mV N 2W 11.003 MATH mem = mem - m[1] 11.004 ACC 100 1.000mV .005% .0035/ 11.005 MEME 11.006 MEMC mV 0.1% 1U #! Test Tol 1.001, Sys Tol 0.000107, TUR 9355.140 (>= 4.00). 12.001 DISP Set the UUT front panel controls as follows: 12.001 DISP 100µV/DIV .................................... 0 12.002 END